Characteristics
General
Number of Channels - (all channels are acquired including clocks).TLA-5201B: 34 channels (2 are clock channels).TLA-5201B: 68 channels (4 are clock channels).TLA-5203B: 102 channels (4 are clock and 2 are qualifier channels).TLA-5204B: 136 channels (4 are clock and 4 are qualifier channels).
Time Stamp - 51-Bits at 125 ps resolution (3.25 days duration).
Clocking/Acquisition Modes - Internal, internal 2X, internal 4X, external, external 2X, source synchronous. 125 ps (8 GHz) MagniVu™ high-speed timing is available simultaneous with all modes.
Input Characteristics (with P64xx probes)
Capacitive Loading - <0.7 pF typical data/clock (P6419).1.4 pF typical data; 2 pF typical clock (P6418).2 pF typical data/clock (P6417, P6434).
Threshold Selection Range - From -2.0 V to +4.5 V in 5 mV increments.Threshold presets include TTL (1.5 V), CMOS (1.65 V), ECL (–1.3 V), PECL (3.7 V), LVPECL (2.0 V), LVCMOS 1.5 V (0.75 V), LVCMOS 1.8 V (0.9 V), LVCMOS 2.5 V (1.25 V), LVCMOS 3.3 V (1.65 V), LVDS (0 V), and user-defined.
Threshold Selection Channel Granularity -Separate selection for each of the clock/qualifier channels and one per group of 16 data channels.
Threshold Accuracy (including probe) -±(100mV).
Input Voltage Range - Operating: -2.5 V to 5.0 V.Nondestructive: ±15 V.
Minimum Input Signal Swing - ±250mV (P6417, P6418, P6419).±300mV (P6434).
Input Signal Minimum Slew Rate - 200 mV/ns typical.
State Acquisition Characteristics
Maximum State Clock Rate - 235 MHz.
Maximum State Data Rate - 470 Mb/s.
State Record Length with Timestamps (half/full channels) - 4/2 Mb , 16/8 Mb, 64/32 Mb.
Setup and Hold Time Selection Range - 16 ns range that may be shifted towards the setup region by 0 ns [+8, -8] ns, 4 ns [+12, -4] ns, or 8 ns [+16, 0] ns.
Setup-and-hold Window - All Channels: 1.5 ns typical.
Minimum Clock Pulse Width - 1.5 ns (P6434).1.25 ns (P6417, P6418, P6419).
Demux Channel Selection - Channels can be demultiplexed to other channels through user interface with 8 channel granularity.
Timing Acquisition Characteristics
MagniVu™ Timing Resolution - 125 ps (8 GHz).Storage rate adjustable to 250 ps, 500 ps, 1 ns, and 2 ns.
MagniVu Timing Record Length - 16 Kb per channel, with adjustable trigger position.
Timing Resolution (quarter/half/full channels) - 500 ps/1 ns/2 ns to 50 ms.
Timing Record Length (quarter/half/full channels with timestamps and with or without transitional storage) -8/4/2 Mb, 32/16/8 Mb, 128/64/32 Mb per channel.
Timing Record Length with Glitch Storage Enabled - Half of main record length.
Channel-to-channel Skew - 1 ns (900 ps typical).
Minimum Recognizable Pulse/Glitch Width (single channel) - 1 ns (P6417, P6418, P6419) 1.25 ns (P6434).
Minimum Detectable Setup/Hold Violation - 250 ps.
Minimum Recognizable Multi-channel Trigger Event - Sample period + channel-to-channel skew.
Trigger Characteristics
Independent Trigger States - 16.
Maximum Independent If/then Clauses per State -16.
Maximum Number of Events per If/then Clause -8.
Maximum Number of Actions per If/then Clause -8.
Maximum Number of Trigger Events - 18 (2 counter/timers plus any 16 other resources).
Number of Word Recognizers - 16.
Number of Transition Recognizers - 16.
Number of Range Recognizers - 4.
Number of Counter/Timers - 2.
Trigger Event Types - Word, group, channel, transition, range, anything, counter value, timer value, signal, glitch, setup-and-hold violation, snapshot.
Trigger Action Types - Trigger main, trigger MagniVu™, store, don't store, start store, stop store, increment counter, decrement counter, reset counter, start timer, stop timer, reset timer, snapshot current sample, goto state, set/clear signal, do nothing.
Trigger Sequence Rate - DC to 500 MHz (2 ns).
Counter/Timer Range - 51 Bits each (>50 days at 2 ns).
Counter Rate - DC to 500 MHz (2 ns).
Timer Clock Rate - 500 MHz (2 ns).
Counter/Timer Latency - 2 ns.
Range Recognizers - Double bounded (can be as wide as any group, must be grouped according to specified order of significance).
Setup-and-hold Violation Recognizer Setup Time Range - From 8 ns before to 7.5 ns after clock edge in 125 ps increments.
Setup-and-hold Violation Recognizer Hold Time Range - From 7.5 ns before to 8 ns after clock edge in 125 ps increments.
Trigger Position - Any data sample.
MagniVu™ Trigger Position - MagniVu position can be set from 0% to 60% centered around the MagniVu trigger.
Storage Control (data qualification) - Global (conditional), by state (start/stop), block, by trigger action, or transitional. Force main prefill selection available.
iView™ (Integrated View) Capability
Overview - The iView capability seamlessly integrates and automatically time-correlates data from the logic analyzer and oscilloscope, so you can transfer analog waveforms from the oscilloscope to the logic analyzer display with the click of a mouse. View time-correlated analog and digital signals side-by-side and pinpoint the source of elusive glitches and other problems in moments.
Number of TDS oscilloscopes that can be connected to a TLA- system - 1.
External Oscilloscopes Supported - For a complete list of currently supported TDS oscilloscopes, please visit our website http://www.tektronix.com/iview.
TLA- Connections - USB, Trigger In, Trigger Out, Clock Out.
TDS Connections - GPIB, Trigger In, Trigger Out, Clock In (when available).
Setup - iView external oscilloscope wizard automates setup.
Data Correlation - After TDS oscilloscope acquisition is complete, data is automatically transferred to the TLA- and time correlated with the TLA- acquisition data.
Deskew - TDS and TLA- data is automatically deskewed and time correlated when using the iView external oscilloscope cable.
iView External Oscilloscope Cable Length - 2 m.
PC Characteristics
Operating System - Microsoft Windows XP Professional with Multi-Lingual User Interface Pack.
Processor - Intel Celeron 2.93 GHz.
Chipset - Intel 945.
DRAM - 512 MB DDR2.
Sound - 16-Bit I/O and Mic-In port.
Hard Drive - 80 GB.
Optical Drive - Internal 24/10/24 CD-RW/DVD-RO.
Floppy Drive - Built-in 3.5 in. 1.44 MB drive.
Integral Controls
Front-Panel Display - Size: 10.4 in. (26.4 cm) diagonal.Type: Active-matrix color TFT LCD with backlight.Resolution: 1024?768.Colors: 256 K.
Simultaneous Display Capability - The front-panel and secondary displays can be operated simultaneously using the same resolution. The secondary external display can be used simultaneously using an independent resolution.
Front-panel Controls - Special function knobs for instrument control and mini-QWERTY keypad.
External Peripheral Interfaces
External Display Port Type - Two female DB15 SVGA.
External Display Resolution - Up to 1600?1200 non-interlaced at 16.8 M colors.
LAN Port Type - 1000 Base-T, RJ-45.
External Keyboard Port Type - PS2 mini-DIN.
External Mouse Port Type - PS2 mini-DIN.
Parallel Interface Port Type - Female DB25.
Parallel Interface Modes - Centronics mode, EPP (Extended Parallel Port), ECP (Microsoft high-speed mode).
Serial Interface Port Type - Male DB9.
Audio Out Port Type - Stereo minijack.
Mic-In Port Type - Minijack.
USB Port - Four USB 2.0.
Symbolic Support
Number of Symbols/Ranges - Unlimited (limited only by amount of virtual memory available on TLA-).
Object File Formats Supported - IEEE695, OMF 51, OMF 86, OMF 166, OMF 286, OMF 386, COFF, Elf/Dwarf 1 and 2, Elf/Stabs, TSF (TSF is a generic ASCII file format documented in the TLA- user manual). If a format is not listed, please contact your local Tektronix representative.
External Instrumentation Interfaces
System Trigger Output - 50 ? BNC type connector.
System Trigger Input - 50 ? BNC type connector.
External Signal Output - 50 ? BNC type connector.
External Signal Input - 50 ? BNC type connector.
Power
Voltage range/frequency - 90-240 VAC at 47-63 Hz.
Input current - 5 A maximum at 90 VAC.
Power consumption - 300 W maximum.
Physical Characteristics
TLA-5000B |
Dimensions |
mm |
in. |
Weight |
kg |
lb. |
Height |
285 |
11.2 |
Width |
438 |
17.5 |
Depth |
288 |
11.35 |
Net (w/o probes) |
12 |
26 |
Shipping (typical) |
18.5 |
41 |
Environmental
Temperature - Operating: +5 °C to +50 °C.Nonoperating: -20 °C to +60 °C.
Humidity - 20% to 80%.Operating: 20% to 80% relative humidity (29 °C maximum wet bulb temperature).Nonoperating: 8% to 80% (29 °C maximum wet bulb temperature).
Altitude - Operating: -1,000 ft. to 10,000 ft. (-305 meters to 3,050 meters).
Safety - UL3111-1, CSA1010.1, EN61010-1, IEC61010-1.